// 500+ Brands

// Authorized Distributor of Over 100 Brands

ACI QTS-61120X-E000000

SKU: CS6287265MPN: QTS-61120X-E000000Alternate Part #: 132700

Toxic Gas Sensor, Chlorine, 0-3 PPM, 4-20mA Output, loop powered, Industrial NEMA 4X, Aluminum Sensor Housing

Contact us for pricing and availability

Product Details

Toxic Gas Sensor, Chlorine, 0-3 PPM, 4-20mA Output, loop powered, Industrial NEMA 4X, Aluminum Sensor Housing

Product Configuration

Gas Type Chlorine
Enclosure NEMA 4X, 7, 9
Configuration Industrial Wall Mount - Space

Product Family Specifications

Accuracy +/- 2.5% of Reading
Sensor Life 2 to 3 years, typical
Sensor Type Electrochemical, non-consuming
Repeatability +/- 1% of Reading
Response Time T90: \< 60 Seconds, typical
Output Signals 4-20 mA (2-Wire, Loop Powered)
Product Weight Industrial “E”/“S” Enclosures: 5.00 lbs (2.27kg) Commercial “C”/“D”/“N” Enclosures: 2.00 lbs (0.91kg)
Unit Shelf Life 6 Months from date of purchase (Must be installed and operational)
Agency Approvals RoHS, CSA-4828-02 (CSA File #: 088890 0 000 for Industrial Enclosure Options Only)
Factory Set Range See Gas Sensor Selection \& Specification Table on back of data sheet
Product Dimensions Commercial “C” Enclosure: 6.60” (167.64 mm) x 4.23” (107.44 mm) x 2.60” (66.04 mm) Commercial “D”\& “N” Enclosure: 6.60” (167.64 mm) x 4.23” (107.44 mm) x 3.70” (93.98 mm) Industrial “E”\& “S” Enclosure: 10.70” (271.78 mm) x 5.71” (145.04 mm) x 4.25” (107.95 mm)
RFI/EMI Protection 4.0 Watts @ 1m (3.28 ft) Radiated
Operating Environment Indoor Use only
Connections | Wire Size Screw Terminal Blocks, 16 to 24 AWG (0.51 to 1.30 mm) Shielded Twisted Pair
Maximum Load (@ 24 VDC) 600 Ohms
Recommended Maintenance Accuracy \& Bump test every 3 to 6 months or as required by Code | Replace sensor as need
Operating Humidity Range 0 to 90% RH, non-condensing (Continuous); 0 to 95% RH, non-condensing (Intermittent)
Operating Temperature Range See Gas Sensor Selection \& Specification Table on back of data sheet
Supply Voltage / Protection +24 VDC nominal (+12 to 36 VDC Load Dependent) | 0.5A Socketed Pico Fuse
Terminal Block Torque Rating 0.37 ft-lb (0.502 Nm) Nominal
Operating Atmospheric Pressure 14.696 psi (1.0132 bar) +/- 10% Note: When installed @ > 3000’ above sea level, the gas transmitters must be veried for accuracy \& re-calibrated as needed after installation
Storage Temperature | RH Range 32 to 68ºF (0 to 20ºC) | 15 to 90% RH, Non-Condensing
Coverage Area / Mounting Height See Gas Sensor Selection \& Specification Table on back of data sheet
Supply Current / Power Consumption 25 mA maximum | 0.9 Watts maximum
Intrinsic Safety Barrier Requirement 28.5 VDC maximum and 300 minimum load Note: Requirements to meet when installed with CSA Approved Intrinsic Safety Barriers in Class A, B, C \& D Certified environments
Enclosure Specifications (Type, Material Type, Flammability, NEMA/IP Rating, Explosion Proof) Commercial “C”: General Purpose (Wall Mount); ABS Plastic, UL94 V-0; NEMA 1 (IP10) Commercial “D”: General Purpose (Duct Mount); ABS Plastic, UL94 V-0; NEMA 1 (IP10) Commercial “N”: Weatherproof (Wall Mount); ABS Plastic; UL94 V-0; NEMA 4X (IP56) Industrial “E”: Industrial Connection Head; Aluminum (Epoxy Coated); NEMA 4X (IP56) and Enclosure meets Type 7 \& 9 Explosion Proof Class 1 Div. 1, Groups B, C, D also Group A; CSA, FM (Factory Mutual), EX (Explosion Proof), UL listed Industrial “S”: Industrial Connection Head; 316 Stainless Steel; NEMA 4X (IP56) and Enclosure meets Type 7 \& 9 Explosion Proof Class 1 Div. 1, Groups B, C, D also Group A; CSA, FM (Factory Mutual), EX (Explosion Proof), UL listed Note for E \& S enclosure: The Cl2, ClO2, HCl, SO2, NO2, NH3, and O3 sensor assemblies are NOT explosion-proof rated

Specifications

SKUCS6287265
MPNQTS-61120X-E000000
Alternate Part #132700